Opto22 Board

  • Opto22 modules act like relays and optically isolate higher voltage input & output signals from the control board to protect its circuitry
  • Control Board connects to the Opto22 board via a 50-pin ribbon cable
  • Colors indicate relay type
  • For output relays, LED indicates that the control board is driving the relay on, not that output voltage is present. LED will still light if the fuse is blown
  • For input relay, LED indicates that the sensor is active, but not that the control board recognizes input

Master Power & Power Control

  • Power to all output relays except gasket charge is routed through these two relays
    • Master Power is normally open
    • Power Control is normally closed
  • Master Power is turned on and off by the ON / OFF switch on the touchpad
  • A blown fuse on any output relay will likely cause blown fuses on both these relays also
  • Relay slot number correspond to schematic slot numbers