Note
The following instructions assume a PSS5 control system
on a PRIMUS sterilizer or a retrofitted PSS5 control kit.

Primary Settings

Go into MENU while in operate mode. If the unit has a vertical touchpad, press the MENU key. If the sterilizer has a horizontal touchpad without a MENU key, touch STER TEMP or STER TIME to get to the MENU.

Note
The following setpoints are needed to run a vacuum cycle

Set the cycle name to VACUUM

Set the number of PRE-VACS at 3 (3 pressure and vacuum pulses) (A minimum of 1 and a maximum of 6 PRE-VACS are allowed with 3 PRE-VACS being standard)

Set the number of POSTVACS at 1 (This enables a post sterilization dry phase)

Set the POSTVACS setpoint at -10.0 (10 inches of mercury vacuum)

If running a 270°F – 273°F (132.2°C) sterilize cycle, use a PRECHARGE PRESS of 20 psig

If running a 250°F (121.1°C) sterilize cycle, use a PRECHARGE PRESS of 10 psig

Typically, set a PREVAC VAC TIME of 1:30

VAC TIME will only occur on the third through sixth PRE-VACS

Set a PREVAC VAC POINT of -20.0 (20 inches of mercury vacuum)

Set a PURGE of 1:30

Refer to the printout below for rough guidelines on other settings.

Depending on what is being sterilized, each end user facility must establish the Sterilize Time and Temperature. ADD PHOTOS


Have a question or having difficulties? Talk to Dave Schall, our PRIMUS sterilizer expert who is here to assist you anytime at 877.679.7800 extension 1212 or email at dschall@spire-is.com.

Visit us at spire-is.com

Back to Blog