Gasket Bleed Valve
A good starting place for adjustment of the gasket bleed needle valve is open 1/8 turn. This valve needs to be just barely open to accomplish its intended function.
Chamber Drain Bleed (Examples)
Model | Chamber Size | Minimum Valve Settings |
AA | 16x16x26 | 1/4 to 3/8 turn |
A | 20x20x38 | 3/8 turn |
B | 26x26x39 | 1/2 turn |
C | 26x36x49 | 1/2 turn |
G | 26x36x60 | 3/4 turn |
H | 36x42x60 | 3/4 turn |