Follow the instructions below to view sensor temperatures and pressures.
Steps
- Remove power to the controls by turning off the breaker switch located on the outside of the control panel
- Open the control panel door and toggle the EPROM switch (located under “CAUTION” on the control board) from OPER to CAL
- Restore power to the controls with the breaker switch
- The display will read “HIT ANY BUTTON”
- Using the [arrow] keys, navigate to “SENSOR DIAGS”
- Press [ENTER]
- You can now navigate between the Chamber, Jacket, and Pressure readings using the [arrow] keys
- When done, power OFF the controls again before toggling the EPROM switch back from CAL to OPER